IcanClave
Icanclave T
T229 | T245
IcanClave T229 and T245 are autoclaves in complete compliance with European StandardEN13060, they have bigger chamber, provide sterilization solution for bigger size instruments or more loads.
Standard pre-programmed cycles: Solid 134℃, Wrapped 134℃, Wrapped 121℃, Prion 134℃, B&D Test, Helix Test, Vacuum Test.
Option cycles: Liquid 121℃, Dry, Customer Define
Steam generator: Built-in steam generator with 2200W rated power.
Water Reservoir: One distilled water reservoir 12L and one wastewater reservoir.
Door system: Vacuum-assisted electronic motorized door system.
Water Quality monitoring: Cycle is not available to start if the water quality worse than the value be set.
LCD Display: Displaying for cycle process, time, temperature, pressure, process alert and error codes.
User friendly interface: Easy to operate.
Warning system: Warning of not enough water in distiller water reservoir. Warning of wastewater tank be full. Warning of door unlock. Warning of printer out of paper.
Atmospheric pressure identification system: It can automatically identify the local atmospheric pressure.
Safety system: Over pressure safety valve. Over temperature switch. Safety lock system preventing improper door lock. Circuit breakers protecting the machine when it has short-circuited or over current. Automatic warning for maintenance.
Extra functions: Option to be connected directly with Water Demineralizer Listo-23.
Printer (option): an internal printer is optional for all ICANCLAVE D autoclaves.
USB port: it is used for connecting a USB disc, all sterilization data are automatically output in the USB disc and can be ready directly in any PC and stored electronically.
Internal Memory: the last 10000+ cycles can be automatically stored in the autoclave system, which can be printed out at any time.
Order Code | IcanClave T229 | IcanClave T245 |
---|---|---|
Model Number | STE-29-T | STE-45-T |
Chamber Dimension | Ø319×417 mm | Ø319×617 mm |
Chamber Volume | appr. 29L | appr. 45L |
Material of Trays | Stainless steel | Stainless steel |
Voltage. Freq. | 220/110V, 50/60Hz | 220/110V, 50/60Hz |
Power | 2300W | 2800W |
Overall Dimension(WxHxD) | 640x560x640 mm | 640x560x840 mm |
Weight | 91kg | 103kg |
Program | Temp./Pressure | Steril. Time | Total Time (min) | |
---|---|---|---|---|
T229 | T245 | |||
Solid 134 | 134℃/210Kpa | 3.5 min | 30-45 | 40-50 |
WRAPPED 134 | 134℃/210Kpa | 4 min | 45-60 | 55-75 |
WRAPPED 121 | 121℃/110Kpa | 20 min | 50-65 | 60-80 |
PRION 134 | 134℃/210Kpa | 18 min | 55-70 | 65-85 |
B&D TEST | 134 | 3.5 min | 35-40 | 50-60 |
HELIX TEST | 134 | 3.5 min | 35-40 | 50-60 |
VACUUM TEST | – | – | 20-25 | 22-27 |
* The cycle time is based on different loading and with/without pre-heating.
* Sterilization time and Drying time can be adjusted.
93/42/EEC Directive on Medical Devices.
Pressure Equipment Directive 2014/68/EU.
EN 13060 Small Steam Sterilizers.
EN 61326, EN 61000 Electromagnetic Compatibility.
EN 61010-1, UL 61010-1Safety requirements.
EN 61010-2-040, IEC 61010-2-040.
RoHS Directive 2011/65/EU.