ClaveMed

Icanclave K

K208 / K212 / K218 / K223

They are full automatic “Class N” autoclaves as same shape as ICANCLAVE D, but with non-vacuum. A independent steam generator was built-in, worked by dynamic pulsed air removal for uniform temperature and full steam distribution.

Overview

They are automatic “Class N” autoclaves as same shape as ICANCLAVE D, but with non-vacuum. An independent steam generator was built-in, worked by dynamic pulsed air removal for uniform temperature and full steam distribution.

Excellent Performance

  • Full automatic table top autoclave with pre-set programs.
  • Worked non- vacuum
  • In built independent rapid steam generator ensures quick sterilization cycles.
  • Rapid post sterilization vacuum drying.
  • With one flash water tank and used one water tank.
  • LCD screen displaying time, temperature and pressure, process alert and instrument
  • User friendly interface.
  • Self-initialization when work in different altitude.
  • Multilanguage menu: Chinese, English, Spanish, German, Lithuanian, Polish, French, Czech, Hungarian, Romanian, Dutch, Latvian, Italian, Portuguese, Russian, Croatian.

Safety and Monitoring

  • Double safety protection system prevents the cycle from starting if the door is not correctly locked.
  • This system also prevents the door from opening if the pressure inside the chamber is not equal to atmospheric pressure outside the chamber.
  • Pressure safety valve prevents over pressure in the chamber and steam generator.
  • The power is automatically cut if a short circuit or electrical fault occurs.
  • Able to detect and identify the exact cause of any problem, and to issue a specific error code to the operator.
  • Water level main switch in the main tank controlling the min and max water level.
  • Automatic warning for maintenance.

Documentation

  • Printer (option): an external printer is optional for all “icanclave”D autoclaves.
  • USB port: it is used for connecting a USB stick, all sterilization data are automatically written in the USB stick, and can be ready directly in any PC and stored electronically.
  • Internal Memory: the last 20 cycles can be automatically stored in the autoclave system, which can be printed out at any time.
Order Code ICANCLAVE K208 ICANCLAVE K212 ICANCLAVE K218 ICANCLAVE K223
Model Number STE-8-D STE-12-D STE-18-D STE-23-D
Chamber Dimension Ø170×320 mm Ø200×360 mm Ø247×350 mm Ø247×450 mm
Chamber Volume 8 L 12 L 18 L 23 L
Number of Trays 2/ Max. 3 2/ Max. 3 3/ Max. 4 3/ Max. 4
Tray size 130×265 mm 156×295 mm 192×285 mm 192×380 mm
Voltage, Freq. 220/110V, 50/60Hz 220/110V, 50/60Hz 220/110v, 50/60Hz 220/110v, 50/60Hz
Power 1900 w 1900 w 1900 w 2100 w
Overall Dimension (WxHxD) 420x370x525 mm 420x370x595 mm 490x455x600 mm 490x455x690 mm
Weight 35 kgs 38 kgs 45 kgs 50 kgs
Program Temp. (°C) Pressure (Mpa) Sterilization time (min.)
SOLID 134 210 4
121 110 20
LIQUID 134 210 10
121 110 30
WRAPPED 134 210 10
121 110 30
TEXTILE 134 210 10
121 110 30
PRION 134 210 18
DRYING