Icanclave E1
It is a semi-automatic “Class N” automatic autoclave 16L, with two programs 121°C and 134 °C. Sterilization time and drying time can be adjustable. Drying and drying plus function can be optional for this series.
It is a semi-automatic “Class N” autoclave 16L, worked without steam generator, with two programs 121°C and 134 °C. It does not have water tank, the water has to been filled inside of the chamber before sterilization. It is a most economical steam sterilizer for instruments of the clinics, beauty salon or tattoo shops.
- E118, as the basic model only offers manual pressure release after the sterilization cycle is over.
- E1318, include automatic water discharging.
- E1418, include drying cycle
- E1518, be equipped with an air compressor that is activated automatically once the cycle is finished improving drying results.
- – Sterilization cycles: 121°C and 134 °C.
- – Sterilization time 121°C: Default 30min, 30~60min adjustable.
- – Sterilization time 134°C: Default 30min, 10~20min adjustable.
- – Drying time: 1~60min adjustable. (Model E1418 and Model E1518 only). (Sólo modelo E1418 y modelo E1518).
- Chamber Size: Ø230×360 mm.
- Voltage: 220V-240V/50Hz, 110V-120V/60Hz
Order Code | ICANCLAVE E1 |
---|---|
Model Number | STE-18-E |
Chamber Dimension | 230×360 mm |
Chamber Volume | 16 L |
Number of Trays | 2 |
Tray size | 285×192 mm |
Voltage, Freq. | 220/110V, 50/60Hz |
Power | 1000 w |
Overall Dimension (WxHxD) | 445x400x605 mm |
Weight | 30 kgs |
Temp. (°C) | Pressure (Mpa) | Sterilization time (min.) |
---|---|---|
134 | 210 | 10 |
121 | 110 | 30 |
- 93/42/EEC Directive on Medical Devices.
- 97/23/EEC Directive on Pressure Equipment.
- EN 13060 Small Steam Sterilizers.
- EN 61326, EN 61000 Electromagnetic Compatibility.
- EN 61010-1, UL 61010-1 Safety requirements.
- EN 61010-2-040, IEC 61010-2-040. Specific requirements for steam sterilizers used for treatment of medical materials.