Standard pre-programmed cycles: Solid 134℃ , Wrapped 134℃, Wrapped 121℃, Prion 134℃, B&D Test, Helix Test, Vacuum Test.
Option cycles: Liquid 121, Dry, Customer define
Steam generator: Built-in steam generator with 2200W rated power.
Water Reservoir: One distilled water reservoir 2.5L and one wastewater reservoir.
LCD Display: Displaying for cycle process, time, temperature, pressure, process alert and error codes.
User friendly interface: Easy to operate.
Warning system: Warning of not enough water in distiller water reservoir. Warning of wastewater tank be full. Warning of door unlock. Warning of printer out of paper.
Atmospheric pressure identification system: It can automatically identify the local atmospheric pressure.
Safety system: Over pressure safety valve. Over temperature switch. Safety lock system preventing improper door lock. Circuit breakers protecting the machine when it has short-circuited or over current. Automatic warning for maintenance.
- Internal printer: It prints the cycle records automatically.
- USB port: it is used for connecting a USB disc, all sterilization data are automatically output in the USB disc and can be ready directly in any PC and stored electronically.
- Internal Memory: the last 10000+ cycles can be automatically stored in the autoclave system, which can be printed out at any time.
Order Code | IcanClave D229 |
---|---|
Chamber Dimension | Ø250x625mm |
Chamber Volume | appr.29L |
Number of Trays | 3 |
Size of Trays(WxH) | Tray #1: 185x570mm Tray #2: 213x570mm Tray #3: 2275x570mm |
Voltage. Freq. | 220/110V, 50/60Hz |
POWER | 2300W |
Overall Dimension (WxHxD) | 490x455x625 mm |
Weight | 65kg |
Program | Temp. /Pressure | Steri. Time | Total Time (min) |
---|---|---|---|
SOLID 134 | 134℃/210Kpa | 4 min | 30-45 |
WRAPPED 134 | 134℃/210Kpa | 4 min | 45-60 |
WRAPPED 121 | 121℃/110Kpa | 20 min | 50-65 |
PRION 134 | 134℃/210Kpa | 18 min | 55-70 |
B&D TEST | 134 | 3.5 min | 35-40 |
HELIX TEST | 134 | 3.5 min | 35-40 |
VACUUM TEST | – | – | 20-25 |
* The cycle time is based on different loading with/without pre-heating.
* Sterilization time and Drying time can be adjusted.
93/42/EEC Directive on Medical Devices.
Pressure Equipment Directive 2014/68/EU.
EN 13060 Small Steam Sterilizers.
EN 61326, EN 61000 Electromagnetic Compatibility.
EN 61010-1, UL 61010-1Safety requirements.
EN 61010-2-040, IEC 61010-2-040.
RoHS Directive 2011/65/EU.