It has same shape and size as IcanClave D208, but with Quick cycle included. Quick cycle take approx. completed time 8 to 15 minutes for unwrapped instruments according to different weight of loading. It is a solution for emergency sterilization using, also an idea for ophthalmic clinic.
Standard pre-programmed cycles: Quick 134℃, Wrapped 134℃, Wrapped 121℃, Prion 134℃, B&D Test, Helix Test, Vacuum Test.
Option cycles: Dry
Steam generator: Built-in steam generator with 2200W rated power.
Water Reservoir: One distilled water reservoir 2.5L and one wastewater reservoir.
LCD Display: Displaying for cycle process, time, temperature, pressure, process alert and error codes.
User friendly interface: Easy to operate.
Warning system: Warning of not enough water in distiller water reservoir. Warning of wastewater tank be full. Warning of door unlock. Warning of printer out of paper.
Atmospheric pressure identification system: It can automatically identify the local atmospheric pressure.
Safety system: Over pressure safety valve. Over temperature switch. Safety lock system preventing improper door lock. Circuit breakers protecting the machine when it has short-circuited or over current. Automatic warning for maintenance.
- Internal printer: it prints cycle records automatically.
- USB port: it is used for connecting a USB disc, all sterilization data are automatically output in the USB disc and can be ready directly in any PC and stored electronically.
- Internal Memory: the last 1000+ cycles can be automatically stored in the autoclave system, which can be printed out at any time.
Order Code | IcanClave D208Q | IcanClave D212Q |
---|---|---|
Chamber Dimension | Ø170x320mm | Ø200x360mm |
Chamber Volume | appr.8L | appr.12L |
Number of Trays | 2/ Max. 3 | 2/ Max. 3 |
Material of Trays | Aluminum | Aluminum |
Size of Trays(WxH) | 130x265mm | 160x315mm |
Voltage. Freq. | 220/110V, 50/60Hz | 220/110V, 50/60Hz |
POWER | 2300W | 2300W |
Overall Dimension(WxHxD) | 420x370x525 mm | 420x370x595 mm |
Weight | 39kgs | 42kgs |
Program | Temp. /Pressure | Steri. Time | Total Time (min) | |
---|---|---|---|---|
D208Q | D212Q | |||
QUICK 134 | 134℃/210Kpa | 4 min | 7-12 | 9-15 |
WRAPPED 134 | 134℃/210Kpa | 4 min | 20-38 | 24-42 |
WRAPPED 121 | 121℃/110Kpa | 20 min | 31-45 | 35-48 |
PRION 134 | 134℃/210Kpa | 18 min | 30-40 | 34-45 |
DRY(option) | – | – | 1-20 | 1-20 |
B&D TEST | 134 | 3.5 min | 10-20 | 15-25 |
HELIX TEST | 134 | 3.5 min | 10-20 | 15-25 |
VACUUM TEST | – | – | 17-20 | 18-21 |
* The cycle time is based on different loading and with/without pre-heating.
* Sterilization time and Drying time can be adjusted.
- 93/42/EEC Directive on Medical Devices.
- Pressure Equipment Directive 2014/68/EU.
- EN 13060 Small Steam Sterilizers.
- EN 61326, EN 61000 Electromagnetic Compatibility.
- EN 61010-1, UL 61010-1Safety requirements.
- EN 61010-2-040, IEC 61010-2-040.