- Standard pre-programmed cycles: Quick S, Standard B, Soft B, Prion 134℃, B&D Test, Helix Test, Vacuum Test.
- Other cycles: Descale (cleaning steam generator and tube).
- Steam generator: Built-in steam generator with 2000W rated power.
- Water Reservoir: External water reservoir and wastewater reservoir.
- Door system: Manual door system.
- Color touch screen: Displaying for cycle process, time, temperature, pressure, process alert messages, error codes, and all detail information.
- User friendly interface: Easy to operate.
- Warning system: Warning of door unlock.
- Atmospheric pressure identification system: It can automatically identify the local atmospheric pressure.
- Safety system: Over pressure safety valve. Over temperature switch. Safety lock system preventing improper door lock. Circuit breakers protecting the machine when it has short-circuited or over current. Automatic warning for maintenance.
- Printer port: it is used for connecting an external printer print the cycle report automatically.
- USB port: it is used for connecting a USB disc, all sterilization data are automatically output in the USB disc and can be ready directly in any PC and stored electronically.
- Internal Memory: the last 9999 cycles can be automatically stored in the autoclave system, which can be printed out at any time.
- Barcode label printer port: it is used for connecting an optional barcode printer prints barcode labels that contain the information related to the sterilization cycle.
Order Code | ClaveMed F03 |
---|---|
Model Number | STE-3-F |
Chamber Dimension (mm) | 192(W) x 45(H) x 285(D) |
Chamber Volume | appr. 3L |
Number of Trays | 1 |
Material of Trays | Aluminum |
Size of Trays (WxH) | 178x285mm |
Voltage. Freq. | 220/110V, 50/60Hz |
Power | 2300W |
Overall Dimension (WxHxD) | 268(W ) x 171(H) x 605 (D) |
Weight | 16kg |
Program | Temp./Pressure | Sterilization Time | Total Time |
---|---|---|---|
Quick S | 134℃/210Kpa | 3.5 min | 7-11 min |
Standard B | 134℃/210Kpa | 4 min | 15-18 min |
Soft B | 121℃/110Kpa | 20 min | 25-30 min |
PRION 134 | 134℃/210Kpa | 18 min | 25-30 min |
B&D TEST | 134℃/210Kpa | 3.5 min | 12-18 min |
HELIX TEST | 134℃/210Kpa | 3.5 min | 12-18 min |
VACUUM TEST | – | – | 15-20 min |
* The cycle time is based on different loading and with/without pre-heating.
* Sterilization time and Drying time can be adjusted.
93/42/EEC Directive on Medical Devices.
Pressure Equipment Directive 2014/68/EU.
EN 13060 Small Steam Sterilizers.
EN 61326, EN 61000 Electromagnetic Compatibility.
EN 61010-1, UL 61010-1Safety requirements.
EN 61010-2-040, IEC 61010-2-040.
RoHS Directive 2011/65/EU.